HP Hewlett Packard Satellite TV System 8517B User Manual

HP 8517B  
S-Parameter Test Set  
Operating and  
Service Manual  
Printed in USA  
12 August 1998  
HP part number 08517-90054  
 
What You’ll Find in This Manual…  
HP 8517B S-Parameter Test Set  
Chapters 1 and 2 Test set description and installation information  
Chapters 3 and 4 Principles of operation and specifications  
Chapters 5, 6, and 7 Troubleshooting, replacement procedures, and replaceable parts  
information  
HP 8517B S-Parameter Test Set Manual iii  
 
   
Warranty  
Certification  
Warranty  
Hewlett-Packard Company certifies that this product met its published  
specifications at the time of shipment from the factory. Hewlett-Packard  
further certifies that its calibration measurements are traceable to the  
United States National Institute of Standards and Technology (NIST,  
formerly NBS), to the extent allowed by the Institute’s calibration facility,  
and to the calibration facilities of other International Standards  
Organization members.  
This Hewlett-Packard system product is warranted against defects in  
materials and workmanship for a period corresponding to the individual  
warranty periods of its component products. Instruments are warranted for a  
period of one year. During the warranty period, Hewlett-Packard Company  
will, at its option, either repair or replace products that prove to be defective.  
Warranty service for products installed by HP and certain other products  
designated by HP will be performed at Buyer’s facility at no charge within  
HP service travel areas. Outside HP service travel areas, warranty service  
will be performed at Buyer’s facility only upon HP’s prior agreement and  
Buyer shall pay HP’s round trip travel expenses. In all other areas, products  
must be returned to a service facility designated by HP.  
For products returned to HP for warranty service, Buyer shall prepay  
shipping charges to HP and HP shall pay shipping charges to return the  
product to Buyer. However, Buyer shall pay all shipping charges, duties, and  
taxes for products returned to HP from another country.  
HP warrants that its software and firmware designated by HP for use with an  
instrument will execute its programming instructions when properly  
installed on that instrument. HP does not warrant that the operation of the  
instrument, or software, or firmware will be uninterrupted or error free.  
LIMITATION OF WARRANTY. The foregoing warranty shall not apply  
to defects resulting from improper or inadequate maintenance by Buyer,  
Buyer-supplied software or interfacing, unauthorized modification or  
misuse, operation outside of the environmental specifications for the  
product, or improper site preparation or maintenance.  
NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HP  
SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OR  
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.  
EXCLUSIVE REMEDIES. THE REMEDIES PROVIDED HEREIN ARE  
BUYER’S SOLE AND EXCLUSIVE REMEDIES. HP SHALL NOT BE  
LIABLE FOR ANY DIRECT, INDIRECT, SPECIAL, INCIDENTAL, OR  
iv HP 8517B S-Parameter Test Set Manual  
 
     
CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,  
TORT, OR ANY OTHER LEGAL THEORY.  
Assistance  
Product maintenance agreements and other customer assistance agreements  
are available for Hewlett-Packard products.  
For assistance, call your local Hewlett-Packard Sales and Service Office  
HP 8517B S-Parameter Test Set Manual  
v
 
 
Service and Support  
Any adjustment, maintenance, or repair of this product must be performed  
by qualified personnel. Contact your customer engineer through your local  
HP Service Center. You can find a list of HP Service Centers on the web at  
http://www.hp.com/go/tmdir.  
If you do not have access to the Internet, one of these HP centers can direct  
you to your nearest HP representative:  
United States:  
Hewlett-Packard Company  
Test and Measurement Call Center  
PO Box 4026  
Englewood, CO 80155-4026  
(800) 452 4844 (toll-free in US)  
Canada:  
Europe:  
Hewlett-Packard Canada Ltd.  
5150 Spectrum Way  
Mississauga, Ontario L4W 5G1  
(905) 206 4725  
Hewlett-Packard European Marketing Centre  
Postbox 999  
1180 AZ Amstelveen  
The Netherlands  
(31 20) 547 9900  
Japan:  
Hewlett-Packard Ltd.  
Measurement Assistance Center  
9-1, Takakura-Cho, Hachioji-Shi  
Tokyo 192, Japan  
(81) 426 56 7832  
(81) 426 56 7840 (FAX)  
Latin America:  
Hewlett-Packard Latin American Region Headquarters  
5200 Blue Lagoon Drive, 9th Floor  
Miami, Florida 33126, U.S.A.  
(305) 267 4245, (305) 267-4220  
(305) 267 4288 (FAX)  
Australia/New Zealand:  
Hewlett-Packard Australia Ltd.  
31-41 Joseph Street  
Blackburn, Victoria 3130  
Australia  
1 800 629 485 (Australia)  
0800 738 378 (New Zealand)  
(61 3) 9210 5489 (FAX)  
Asia-Pacific:  
Hewlett-Packard Asia Pacific Ltd.  
17-21/F Shell Tower, Times Square  
1 Matheson Street, Causeway Bay  
Hong Kong  
(852) 2599 7777  
(852) 2506 9285 (FAX)  
vi HP 8517B S-Parameter Test Set Manual  
 
       
Safety and Regulatory Information  
Review this product and related documentation to familiarize yourself with  
safety markings and instructions before you operate the instrument. This  
product has been designed and tested in accordance with international  
standards.  
WARNING  
CAUTION  
The WARNING notice denotes a hazard. It calls attention to a procedure,  
practice, or the like, that, if not correctly performed or adhered to, could result  
in personal injury. Do not proceed beyond a WARNING notice until the  
indicated conditions are fully understood and met.  
The CAUTION notice denotes a hazard. It calls attention to an operating  
procedure, practice, or the like, which, if not correctly performed or adhered  
to, could result in damage to the product or loss of important data. Do not  
proceed beyond a CAUTION notice until the indicated conditions are fully  
understood and met.  
Instrument Markings  
When you see this symbol on your instrument, you should refer to the instrument’s  
instruction manual for important information.  
!
This symbol indicates hazardous voltages.  
The laser radiation symbol is marked on products that have a laser output.  
This symbol indicates that the instrument requires alternating current (ac) input.  
The CE mark is a registered trademark of the European Community. If it is  
accompanied by a year, it indicates the year the design was proven.  
The CSA mark is a registered trademark of the Canadian Standards Association.  
1SM1-A  
This text indicates that the instrument is an Industrial Scientific and Medical Group 1  
Class A product (CISPER 11, Clause 4).  
This symbol indicates that the power line switch is ON.  
This symbol indicates that the power line switch is OFF or in STANDBY position.  
HP 8517B S-Parameter Test Set Manual vii  
 
 
Safety Earth  
Ground  
This is a Safety Class I product (provided with a protective earthing  
terminal). An uninterruptible safety earth ground must be provided from the  
main power source to the product input wiring terminals, power cord, or  
supplied power cord set. Whenever it is likely that the protection has been  
impaired, the product must be made inoperative and secured against any  
unintended operation.  
Before Applying Power Verify that the product is configured to match the available main power  
source as described in the input power configuration instructions in this  
manual. If this product is to be powered by auto-transformer, make sure the  
common terminal is connected to the neutral (grounded) side of the ac power  
supply.  
viii HP 8517B S-Parameter Test Set Manual  
 
   
Typeface Conventions  
Not all of the following conventions may appear within this manual,  
however, refer to this listing whenever you encounter on of the special font  
characters.  
Italics Used to emphasize important information:  
Use this software only with the HP 8517B Test Set.  
Used for the title of a publication:  
Refer to the HP 8517B S-Parameter Test Set Manual  
Used to indicate a variable:  
Type LOAD BIN filename.  
Instrument Display  
Used to show on-screen prompts and messages that you will see on the  
display of an instrument:  
The HP 8517B will display the message CAL1 SAVED.  
Keycap  
Used for labeled keys on the front panel of an instrument or on a  
computer keyboard:  
Press [Return].  
Softkey  
Used for simulated keys that appear on an instrument display:  
Press {Prior Menu}.  
User Entry  
Used to indicate text that you will enter using the computer keyboard;  
text shown in this typeface must be typed exactly as printed:  
Type LOAD PARMFILE  
Used for examples of programming code:  
#endif // ifndef NO_CLASS  
Path Name  
Used for a subdirectory name or file path:  
Edit the file usr/local/bin/sample.txt  
Computer Display  
Used to show messages, prompts, and window labels that appear on a  
computer monitor:  
The Edit Parameters window will appear on the screen.  
Used for menus, lists, dialog boxes, and button boxes on a computer  
monitor from which you make selections using the mouse or keyboard:  
Double-click EXIT to quit the program.  
HP 8517B S-Parameter Test Set Manual ix  
 
 
x
HP 8517B S-Parameter Test Set Manual  
 
Contents  
Table 1-1. HP 8517B Options and Descriptions . . . . . . . . . . . . . . . . 1-4  
Table 1-2. Test Set Operation Verification . . . . . . . . . . . . . . . . . . . . . 1-5  
Table 1-4. Service and Support Options . . . . . . . . . . . . . . . . . . . . . . 1-9  
Table 1-6. Recommended Test Equipment . . . . . . . . . . . . . . . . . . . . 1-12  
Table 1-7. Maximum Input Power Levels . . . . . . . . . . . . . . . . . . . . 1-13  
HP 8517B S-Parameter Test Set Manual Contents-1  
 
Table 2-1. Accessories in Figure 2-1 . . . . . . . . . . . . . . . . . . . . . . . . 2-3  
Table 3-1. Front-Panel Description . . . . . . . . . . . . . . . . . . . . . . . . . 3-1  
Table 3-2. Rear-Panel Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . 3-2  
Contents-2 HP 8517B S-Parameter Test Set Manual  
 
Figure 3-5. Using a Torque Wrench Correctly . . . . . . . . . . . . . . . . 3-16  
Figure 3-7. Mating the Clamp and Nut Flats . . . . . . . . . . . . . . . . . 3-17  
The S and S Measurements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3  
11  
21  
Figure 5-2. Troubleshooting Flowchart . . . . . . . . . . . . . . . . . . . . . . 5-5  
Table 5-1. Equipment Required, But Not Supplied . . . . . . . . . . . . . . 5-6  
Table 5-2. Power Supply Voltages to A15 . . . . . . . . . . . . . . . . . . . . . 5-7  
Table 5-3. Power Supply Voltages to A27 . . . . . . . . . . . . . . . . . . . . . 5-7  
Figure 5-3. Power Supply Fuses and Test Points . . . . . . . . . . . . . . . 5-8  
Figure 5-4. Instrument HP-IB Switch Setting . . . . . . . . . . . . . . . . . . 5-9  
Figure 5-5. Fuse and Voltage Cam Location . . . . . . . . . . . . . . . . . . . 5-9  
Figure 5-6. The ACTIVE LED Location . . . . . . . . . . . . . . . . . . . . . 5-10  
Figure 5-8. Service Adapter Connections . . . . . . . . . . . . . . . . . . . . 5-12  
Table 5-5. Results Observations . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13  
HP 8517B S-Parameter Test Set Manual Contents-3  
 
Figure 5-9. Diagram of IF Signal Path . . . . . . . . . . . . . . . . . . . . . 5-14  
HP 8517B Option 007 Test Set . . . . . . . . . . . . . . . . . . . . . . . . . 5-20  
HP 8517B Option 007 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-21  
(B indicates Bad; — indicates Good) . . . . . . . . . . . . . . . . . . . . 5-23  
Figure 6-1. ESD-Safe Workstation Configuration . . . . . . . . . . . . . . 6-2  
Table 6-1. Replacement-Procedure Tools Needed . . . . . . . . . . . . . . 6-3  
Contents-4 HP 8517B S-Parameter Test Set Manual  
 
Table 7-1. Reference Designation Abbreviations . . . . . . . . . . . . . . 7-2  
Table 7-2. Manufacturer Names and Addresses . . . . . . . . . . . . . . . . 7-3  
Table 7-3. Standard Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4  
Figure 7-1. Power Cable and Plug Part Numbers . . . . . . . . . . . . . . 7-11  
Table 7-5. Instrument Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12  
7-14  
HP 8517B Test Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-15  
HP 8517B Test Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16  
HP 8517B Test Set . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17  
Table 7-13. Miscellaneous Parts, Motherboard . . . . . . . . . . . . . . . 7-25  
HP 8517B S-Parameter Test Set Manual Contents-5  
 
Contents-6 HP 8517B S-Parameter Test Set Manual  
 
1
Getting Started  
This is the operating and service manual for the HP 8517B S-parameter test  
set. It is to be used in conjunction with the HP 8510C Network Analyzer  
Operating and Service Manual. Together, these manuals provide  
information needed to configure the system and make measurements.  
How to Use This Manual  
Step 1.  
Review Chapter 1, “Getting Started” and Chapter 2, “Installation” of this manual to  
learn about:  
Using this test set  
Using options of the test set  
Preparing the site for operation  
Understanding safety considerations  
Unpacking the instrument and checking it for shipment damages  
Configuring the HP 8517B test set with the HP 8510C Network Analyzer  
Step 2.  
Step 3.  
Insert the contents of this manual into the HP 8510C Test Sets and Accessories  
binder behind the tab labeled “Test Sets.”  
Review the Connector Care Quick Reference Card (HP part number 08510-90360)  
supplied with the HP 8510C manuals set.  
Refer to the calibration kit documentation for complete information about care,  
cleaning, gaging, and connecting precision devices. Knowledge and application of  
proper connector care is essential to achieving good connections and maintaining  
maximum performance quality with your precision calibration devices.  
Step 4.  
Step 5.  
Read Chapter 3, “Operation,” to learn about test set front- and rear-panel features.  
This chapter also has information about controlling multiple test sets, measuring  
high power devices, using the anti-rotation clamps on the cables, and connecting  
devices to the test set.  
Refer to Chapters 4 through 7 for reference information about specifications,  
troubleshooting, replacement procedures, and replacement parts information.  
HP 8517B S-Parameter Test Set Manual 1-1  
 
   
Getting Started  
Test Set Description  
Test Set Description  
The HP 8517B test set configured with an HP 8510B/C network analyzer  
and an HP 8360 series source, creates a system capable of making  
S-parameter measurements from 45 MHz to 50 GHz. The system is  
particularly suited for making two-port device measurements. You can  
measure all four S-parameters without physically reversing the DUT (device  
under test).  
The system is also designed for making measurements on non-reciprocal  
devices or components like transistors, amplifiers or isolators where S  
measurements are required.  
12  
Two directional couplers are used for signal separation. The configuration is  
illustrated in the following figures:  
For active-device measurements, two bias tees apply external DC bias to  
both test port center conductors.  
Figure 1-1 Standard HP 8517B Test Set Block Diagram  
1-2 HP 8517B S-Parameter Test Set Manual  
 
             
Getting Started  
Test Set Description  
Figure 1-2 Option 004, HP 8517B Test Set Block Diagram  
Figure 1-3 Option 007, HP 8517B Test Set Block Diagram  
HP 8517B S-Parameter Test Set Manual 1-3  
 
       
Getting Started  
HP 8517B Test Set Options  
HP 8517B Test Set Options  
Options for the HP 8517B test set are described in Table 1-1, below.  
Table 1-1 HP 8517B Options and Descriptions  
Options  
HP 8517B Test Set Option Description  
Option 001  
Adds IF switching capability for up to four test set connections to the HP 8510C Network  
Analyzer.  
The test set used is selected from the network analyzer. Refer to “Controlling Multiple Test  
Sets” in Chapter 3, “Operation” of this manual for more information.  
Option 002  
Option 004  
Deletes the programmable attenuators and bias tees.  
Note: If bias is required but attenuation is not, the bias can be applied externally by using an  
HP 11612B bias tee.  
Moves Port 2 attenuator in front of the b2 sampler, allowing devices with output power up to 30  
dBm (1 W) to be measured. Adds a 3 dB attenuator which protects the switch splitter from high  
power into Port 2.  
Option 007  
Option 908  
Adds five amplifiers, two 10 dB attenuators, two 6 dB attenuators, and two tapered attenuators.  
These components increase the available power and dynamic range of the test set.  
Supplies the hardware required for rack mounting the test set when its handles are removed.  
Refer to Chapter 2, “Installation” for more information.  
Option 910  
Option 913  
Adds a duplicate copy of this manual to the shipment.  
Supplies the hardware required for rack mounting the test set with its handles attached. Refer  
to Chapter 2, “Installation” for more information.  
Option 002 “plus” 007  
Option 004 “plus” 007  
Removes the bias tee, and retains the high dynamic range capability in the test set.  
Adds high-dynamic range and high power measurement capability to Port 2.  
1-4 HP 8517B S-Parameter Test Set Manual  
 
       
Getting Started  
Verifying Test Set Operation  
Verifying Test Set Operation  
The test set is designed to operate with an HP 8510C network analyzer. You  
can verify its operation by:  
Table 1-2 Test Set Operation Verification  
Checking System Operation  
Checking Specifications  
Perform a system calibration as described in the HP 8510C Operating and  
Programming Manual. A successful calibration indicates that the system, and  
therefore the test set, are operating properly.  
Specifications for the test set, and the system, can be determined by running  
the specification and performance verification software described in the HP  
8510C On-Site Service Manual, Chapter 8, “Specification and Performance  
Verification”. Additional mechanical specifications and supplemental  
characteristics are in Chapter 4, “Specifications” of this manual.  
Troubleshooting the Test Set  
To troubleshoot the test set, refer to the HP 8510C On-Site Service Manual.  
Use the information in that manual to determine if the test set is at fault. If the  
test set is at fault, refer to Chapter 5, “Troubleshooting the Test Set” in this  
manual to isolate the trouble.  
HP 8517B S-Parameter Test Set Manual 1-5  
 
                 
Getting Started  
Maintaining Measurement Accuracy  
Maintaining Measurement Accuracy  
Precision measurements rely on a precision calibration of the network  
analyzer. As a general rule, the shorter the time lapse between a calibration  
and a device measurement, the more precise the measurement results will be  
(within the limitations of your system).  
Hewlett-Packard recommends calibrating your system every few hours, or at  
least re-verifying your system’s calibration this frequently. Doing so helps  
retain measurement precision.  
The frequency of calibrations is determined by both the location of the  
system and the ambient temperature stability in its operating area.  
1-6 HP 8517B S-Parameter Test Set Manual  
 
     
Getting Started  
Instruments Supported by This Manual  
Instruments Supported by This Manual  
A serial number label is attached to the rear panel of the test set. There are  
two parts to the serial number: a prefix, and a suffix. See Figure 1-4, and  
read the following descriptions:  
The prefix includes the first four digits of the serial number, plus the  
letter. The contents of this manual apply directly to test sets with the  
same serial number prefix as the one(s) on the title page.  
The suffix includes the last five digits of the serial number, which is  
sequential and unique to each test set. You will need this part of the  
number to report test-set problems to an HP sales and service engineer.  
serial.tif  
Figure 1-4 Instrument Serial-Number Label Example  
HP 8517B S-Parameter Test Set Manual 1-7  
 
     
Getting Started  
Instrument Firmware Compatibility  
Instrument Firmware Compatibility  
To use the HP 8517B test set at frequencies up to 50 GHz, use an  
HP 83650B or HP 83651B source.  
If your network analyzer or source do not meet the required frequency  
ranges for your measurement needs, you need to upgrade your system.  
Please contact an HP Sales and Service representative for information. These  
offices are listed in “Service and Support” on page vi at the front of this  
manual  
Table 1-3 Source Firmware Revision Compatibility Matrix  
Compatible With HP  
8340 Series  
Sources?  
Compatible With  
HP 8350 Series  
Sources?  
Instrument Model  
Number and Name  
Compatible With Model:  
Firmware Revision  
HP 8517B Test Set  
HP 8510B Network Analyzer  
HP 8510C Network Analyzer  
B.06.00 and higher  
C.06.00 and higher  
Yes, up to the higher  
frequency limits.  
Yes, up to the higher  
frequency limits.  
HP 8517B,  
HP 8510C Network Analyzer  
7.0 or higher, only  
Yes, up to the higher  
frequency limits.  
No, not compatible  
Option 007 Test Set  
1-8 HP 8517B S-Parameter Test Set Manual  
 
           
Getting Started  
Service and Support Options Available  
Service and Support Options Available  
A variety of service and support products are available for coverage of  
repair, calibration, and verification issues. Contact your HP sales and service  
engineer for details. Their office addresses and phone numbers are listed in  
the front of this document.  
The purchase of an HP 8517B includes a one year on-site service warranty.  
In the event of failure, Hewlett-Packard provides service for the system.  
NOTE  
System installation is not included.  
Table 1-4 Service and Support Options  
Option Number  
Option Description  
Option W30  
Adds a three year customer return-repair coverage warranty to the instrument.  
Customers may return the instrument to HP within that three year period for repair.  
Option W31  
Adds a three year on-site repair coverage warranty for next-day on-site repair of the  
instrument. Customers may return the instrument to HP within that three year period for  
repair and get next-day service on their instrument.  
Option 1BN  
Option 1BP  
Adds to the instrument a MIL-STD 45662A Certificate of Calibration. This option must be  
ordered when the instrument order is placed.  
Adds to the instrument a MIL-STD 45662A Certificate of Calibration and the  
corresponding calibration data. This option must be ordered when the instrument order is  
placed.  
Option UK6  
Adds a certificate of calibration (rather than the MIL-STD certificate) and the  
corresponding calibration data to the instrument. This option must be ordered when the  
instrument order is placed.  
HP 8517B S-Parameter Test Set Manual 1-9  
 
       
Getting Started  
Accessories  
Accessories  
The accessories supplied with the test set, including part numbers, are listed  
of this manual.  
Accessories Available  
Table 1-5 describes accessories available for the HP 8517B test set. For  
additional HP 8510C system accessories information, refer to the HP 8510C  
manual set.  
Table 1-5 HP 8517B Test Set Accessories Available  
Accessory Type  
Accessory Model Number  
Contents or Description  
Calibration Kit1  
HP 85056A 2.4 mm Calibration Kit  
Contains open and short circuits  
Fixed and sliding loads (2)  
2.4 mm to 2.4 mm adapters  
2.4 mm connector tools and gauges  
Verification Kit1  
Cables  
HP 85057S 2.4 mm Verification Kit  
Contains precision airline, mismatched airline,  
20 dB and 40 dB attenuators  
HP 85133C 2.4 mm Test Port Return  
Cable  
Used when measuring a 2.4 mm-ported device connected directly to  
Port 1 of the test set. The test port return cable is connected  
between the device under test and Port 2.  
HP 85133D 2.4 mm Test Port Return  
Cable Set  
Used when measuring a 2.4 mm-ported device connected between  
the cable ends.  
HP 85133E 2.4 mm Flexible Test Port  
Return Cable  
Used when measuring a 2.4 mm-ported device connected directly to  
Port 1 of the test set. The test port return cable is connected  
between the device under test and Port 2.  
HP 85133F 2.4 mm Flexible Test Port  
Return Cable Set  
Used when measuring a 2.4 mm-ported device connected between  
the cable ends.  
HP 85134D 3.5 mm Test Port Return  
Cable Set  
Used when measuring a 3.5 mm device under test connected  
between the cable ends.  
HP 85134E 3.5 mm Flexible Test Port  
Return Cable  
Used when measuring one end of a 3.5 mm device connected  
directly to an HP 85130F adapter at Port 1. The test-port return  
cable is connected between the device and Port 2.  
HP 85134F 3.5 mm Flexible Test Port  
Return Cable Set  
Used when measuring a 3.5 mm device connected between the  
cable ends.  
HP 85135C 7 mm Test Port Return  
Cable  
Used when measuring one end of a 7 mm device connected  
directly to an HP 85130E adapter at Port 1. The test port return  
cable is connected between the device and Port 2.  
Cables (Continued)  
HP 85135D 7 mm Test Port Return  
Cable Set  
Used when measuring a 7 mm device connected between the cable  
ends.  
1-10 HP 8517B S-Parameter Test Set Manual  
 
         
Getting Started  
Accessories  
Table 1-5 HP 8517B Test Set Accessories Available (Continued)  
Accessory Type  
Accessory Model Number  
Contents or Description  
HP 85135E 7 mm Flexible Test Port  
Return Cable  
Used when measuring one end of a 7 mm device connected directly  
to an HP 85130E adapter at Port 1. The test port return cable is  
connected between the device and Port 2.  
HP 85135F 7 mm Flexible Test Port  
Return Cable Set  
Used when measuring a 7 mm device connected between the cable  
ends.  
Adapters  
HP 85130E Special 2.4 mm to 7 mm  
Adapter Set  
Used to convert special 2.4 mm test set ports into a 7 mm connector  
interface (m or f).  
HP 85130F Special 2.4 mm to 3.5 mm  
Adapter Set  
Used to convert special 2.4 mm ports of the test set into a 3.5 mm  
connector interface (m or f).  
HP 85130G Special 2.4 mm to 2.4 mm  
Adapter Set  
Used to convert special 2.4 mm ports of the test set into a standard  
2.4 mm connector interface (m or f). These adapters function as  
“test port savers.”  
HP 11904S 2.4 mm to K-2.9 mm2  
Adapter Kit  
Used to calibrate the test set using 2.4 mm devices, then to change  
the test ports to 2.92 mm and perform fully error corrected  
measurements. The kit contains (2) 2.4 mm to 2.92 mm (m)  
adapters and (2) 2.4 mm to 2.92 (f) adapters.  
Test Fixture Kit  
HP 85041A Transistor Test Fixture Kit  
(TTF)  
A comprehensive measurement system for testing and  
characterizing stripline packaged microwave transistors. Although  
the fixture contains 7 mm connectors and its frequency limit is  
18 GHz, the kit may be adapted for use with the test set. Use  
HP 85135C or HP 85135E cables with the HP 85130E adapter set.  
Please consult your HP Sales and Service representative for  
recommendations.  
1. NIST traceable data and uncertainties are available.  
2. The K-connector is developed and manufactured by the Wiltron Company (Morgan Hill, California).  
NOTE  
For more information about other 2.4 mm adapters, refer to the “2.4 mm  
Adapters and Calibration Accessories,” Operating Note, (HP part number  
11900-90003).  
HP 8517B S-Parameter Test Set Manual 1-11  
 
         
Getting Started  
Recommended Test Equipment  
Recommended Test Equipment  
Additional equipment and accessories required for use with the test set are  
listed in Table 1-6. The table lists which items are required to verify the  
performance of the test set and which are required to operate it. Other  
equipment may be substituted if their specifications meet or exceed those  
listed in the critical specifications column  
Table 1-6 Recommended Test Equipment  
Critical  
Specifications  
Item  
Recommended Model  
Use  
Network Analyzer  
Source  
no substitute  
no substitute  
no substitute  
HP 8510B or HP 8510C  
HP 83651B  
O, P, T  
O, P, T  
N/A  
Controller  
HP 9000 Series 200 or 300 with 3 Mbyte RAM and  
HP BASIC 3.0 or higher or PCC-305 or PC-308  
HP BASIC Controller with 3 Mbyte Basic Language  
Processor RAM  
Disk Drive  
is compatible with the  
controller  
is compatible with the controller  
P
Multimeter  
range: 0 to 50 V  
HP 3456A  
HP 1740A  
T
T
Oscilloscope  
50 MHz bandwidth  
O = Operation; P = Performance Test; T = Troubleshooting  
1-12 HP 8517B S-Parameter Test Set Manual  
 
     
Getting Started  
Operating and Safety Precautions  
Operating and Safety Precautions  
CAUTION  
ESD Sensitive Assemblies  
Test set assemblies are very sensitive to damage from electrostatic discharge.  
They may or may not continue to function if subjected to electrostatic  
discharge. Their reliability, however, will be impaired. Handle the  
instrument devices at static-safe work stations, only.  
Operating Precautions  
Observe the following normal precautions when handling and operating the  
test set:  
Do not exceed the input power levels listed below:  
Table 1-7 Maximum Input Power Levels  
Power Level  
+17 dBm  
Test Port  
Port 1  
+17 dBm  
Port 2  
+30 dBm1  
Port 2 (Option 004, or Option 004 “plus” 007)  
1. This is the maximum input power allowed with HP 8517B Option 004, or  
Option 004 “plus” Option 007, when the attenuator is set to  
attenuation, the maximum input power allowed is +17 dBm.  
20 dB. Without  
Do not exceed +15 dBm into the test set with the source RF input.  
Never apply a dc voltage to the source RF input of the test set.  
Do not torque any connection at the test port connectors to more than  
90 N-cm (8 in-lb). The wrench supplied with your accessory kit is  
calibrated for 90 N-cm (8 in-lb) torque.  
Do not torque any connection to more than 90 N-cm (8 in-lb) at the  
source RF input or on the back of your test set.  
Safety Precautions  
The voltages inside this test set warrant normal caution for operator safety.  
Nevertheless, service should be performed by qualified personnel, only.  
Service strategy, troubleshooting procedures, replaceable parts, and other  
information about the test set are provided in this manual or the HP 8510C  
On-Site Service Manual.  
HP 8517B S-Parameter Test Set Manual 1-13  
 
           
Getting Started  
Operating and Safety Precautions  
1-14 HP 8517B S-Parameter Test Set Manual  
 
2
Installation  
This chapter contains installation information. Topics include initial  
inspection, environmental considerations, test set location when using it with  
the HP 8510C network analyzer, and making connections to the test set.  
Refer to “Installation” in the HP 8510C On-Site Service Manual for  
complete system connection and turn-on instructions.  
Refer to “Packaging the Test Set” for information about shipping the  
instrument.  
HP 8517B S-Parameter Test Set Manual 2-1  
 
 
Installation  
Initial Inspection  
Initial Inspection  
Inspect the shipping container and cushioning materials for damage. If there  
is damage, keep the container until you have checked the contents for  
completeness.  
If shipping materials are damaged, complete the performance tests outlined  
in the HP 8510C On-Site Service Manual. If the test set fails the  
performance tests, or is damaged or defective, keep the shipping materials.  
Notify both the carrier and the nearest Hewlett-Packard Sales and Service  
Office (listed in “Service and Support” at the front of this manual).  
The office can arrange for repair or replacement of the test set without  
waiting for the claim settlement.  
Operating  
Environment  
For the HP 8517B to operate within specifications, the ambient temperature  
must remain between 0° C and +55° C. Keep relative humidity to less than  
95% (at 40° C dry bulb temperature, maximum). The instrument can be  
operated at altitudes up to 4,500 meters (15,000 feet).  
Storing the Test Set  
Accessories Supplied  
Acceptable storage temperatures range from –40° C to +75° C, with relative  
humidity 90% at +65° C (maximum dry bulb temperature) and at altitudes  
up to 15,240 meters (50,000 feet).  
Figure 2-1 illustrates accessories supplied with the test set. If you did not  
receive one of these accessories with the test set, notify your nearest HP  
Sales and Service Office so that the missing parts can be sent to you.  
2-2 HP 8517B S-Parameter Test Set Manual  
 
               
Installation  
Initial Inspection  
Figure 2-1 Accessories Supplied with the HP 8517B Test Set  
Table 2-1 Accessories in Figure 2-1  
(1) Power Cord U.S.A. only  
(6) Female 2.4 mm to Female 2.4 mm Adapter  
(7) 20 mm, 98 N-cm (8 in-lb) Torque Wrench  
(8) 1/2” x 9/16” Open End Wrench  
(2) Test Set Interconnect Cable  
(3) HP-IB Cable  
(4) Specification and Performance Verification Software  
(5) Anti-rotation Clamps  
(9) RF Source Cable, Bench-top Configuration  
(10) RF Source Cable, Rack Configuration  
CAUTION  
CAUTION  
Assemblies in the test set are extremely sensitive to damage by static  
electricity. They may or may not continue to function if subjected to an  
electrostatic discharge. Their reliability, however, will be impaired.  
Always use an anti-static wrist strap when calibrating or verifying the test set  
or while using the test set to measure devices. Never touch test port center  
conductors. Use a wrist strap when connecting the extended center  
conductor of a sliding load termination.  
HP 8517B S-Parameter Test Set Manual 2-3  
 
     
Installation  
Installing the Test Set Into a System Rack  
Installing the Test Set Into a System Rack  
The HP 85043C system rack is the recommended model to use. To  
rack-mount the test set into a system configured with the HP 8510C, refer to  
the “Installation” chapter of the HP 8510C On-Site Service Manual and to  
the HP 85043C System Rack Manual.  
Installing the Test Set  
on a Bench  
When installing the test set for use on a bench, place it on a grounded  
anti-static work surface to reduce the chance of ESD damage. The antistatic  
surface should extend far enough in front of the test set to provide effective  
protection for the test ports and cable ends. See Figure 2-2.  
A grounding receptacle is provided on the test set as an alternate grounding  
point for your anti-static wrist-strap.  
Figure 2-2 Recommended Static-Free Workstation Configuration  
2-4 HP 8517B S-Parameter Test Set Manual  
 
         
Installation  
Configuring the Test Set in a System  
Configuring the Test Set in a System  
Figure 2-3 illustrates the cable connections for installing an HP 8517B in a  
system. Refer to the information that follows to install the test set.  
Figure 2-3 Configuring an HP 8517B Test Set in a System  
HP 8517B S-Parameter Test Set Manual 2-5  
 
       
Installation  
Configuring the Test Set in a System  
Connecting the System  
Cables  
Table 2-2 Connection Instructions and Connector Descriptions  
Connect System Power  
and Control Cables  
Check the test-set line module (see item (1) in Chapter 3, Figure 3-2 of this manual). It must be set for  
the correct voltage in your region. Refer to Chapter 5, “Troubleshooting” of this manual for information  
about changing the voltage selection or replacing the line fuse.  
After setting or confirming the line-voltage module, connect the test set power cord to an electrical  
outlet.  
Attach the HP 8517B IF-interconnect cable from J11 TEST SET INTERCONNECT (on the test set’s  
rear panel) to J1 TEST SET INTERCONNECT on the rear panel of the HP 85102 IF detector.  
Attach the system bus cable from HP 8517B J12 HP 8510 SYSTEM BUS connector to HP 8510C  
INTERCONNECT on the HP 85101 display/processor.  
The test set IF interconnect cable and the system bus cable transmit control signals between the test  
set and the network analyzer.  
Signal Path Connections  
The IF interconnect cable transmits IF signals from the test set to the HP 85102 IF detector.  
Attach one end of the 2.4 mm RF cable (item 9 or 10, Figure 2-1) to the RF output on the rear panel  
of the HP 83651 Synthesized Sweeper.  
Attach the other end of the RF cable to the RF input on the HP 8517B Test Set.  
Test Port Connectors  
Anti-Rotation Clamps  
Port 1 and Port 2 are male, NMD-2.4 mm connectors, and mate to female 2.4 mm connectors.  
Use the anti-rotation clamps (item 5, Figure 2-1) to stabilize the test port or RF cable at the  
connection, or to stabilize an adapter at the connection.  
Connect test port cables or adapters to the test ports.  
Torque the connections to 90 N-cm (8 in lb).  
Loosen the anti-rotation clamp thumb screw enough to slip the clamp over the cable and up to the  
front panel. The clamp end with the flats should come to rest on the flats of the test port shoulder.  
Finger-tighten the thumb screws to prevent further loosening or tightening of the test port/RF cable  
connection. Refer to Chapter 3, “Operation” of this manual for additional information.  
Replacing O-Rings in  
Anti-Rotation Clamps  
The anti-rotation clamp’s internal O-ring is a field-replaceable part. There is  
no need to disassemble the anti-rotation clamp.  
When the O-ring no longer holds the RF cable securely, replace it by  
following the instructions in Chapter 6, “Replacement Procedures.”  
2-6 HP 8517B S-Parameter Test Set Manual  
 
                     
Installation  
Packaging the Test Set  
Packaging the Test Set  
To repackage the test set, use its original factory packaging.  
Refer to the test set using its complete model and serial numbers in any  
correspondence to an HP Sales and Service Office.  
Containers and materials identical to those used in the original shipment by  
the factory are available through Hewlett-Packard Sales and Service Offices.  
However, commercially available, comparable packaging materials may be  
used, also.  
1. Wrap the test set in heavy paper or anti-static plastic.  
2. If you are shipping the test set to an HP Sales or Service office, complete  
and attach a service tag (located in the HP 8510C manual set).  
3. Use sufficient shock absorbing material on all sides of the test set to  
provide a thick, firm cushion and prevent movement.  
4. Seal the shipping container securely and mark it “FRAGILE.”  
HP 8517B S-Parameter Test Set Manual 2-7  
 
   
Installation  
Packaging the Test Set  
2-8 HP 8517B S-Parameter Test Set Manual  
 
3
Operating the HP 8517B Test Set  
This chapter contains illustrations and descriptions of the HP 8517B front-  
and rear-panel features. Information about operating multiple test sets  
(Option 001) is also provided.  
Front-Panel Features  
Figure 3-1 Front-Panel Features of the HP 8517B Test Set  
Table 3-1 Front-Panel Description  
(1) Line Switch  
This switch turns the test set on or off. When the side of the switch labeled “0” is depressed, the  
test set is OFF; when the side labeled “1” is depressed, the test set is ON.  
(2) Line LED  
This LED is